NEC GammaCompMD QA Manuale Utente Pagina 166

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 165
GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 166
No
Occurrence/Error
Solution
39
When a NEC EA model is
calibrated, which is connected
via DisplayPort , the luminance
is lower than the target
luminance.
Select the "SOUND INPUT" on the OSD Menu.
Touch the DOWN key of the UP-DOWN key and
the INPUT key simultaneously.
When displayed as EDID128, turn the display’s
power OFF and ON again.
40
When a NEC EA model is
calibrated which is connected
via HDMI, calibration is failed.
Set the color format of the graphic card to RGB
[0-255].
Set VIDEO LEVEL of Display to NORMAL.
Vedere la pagina 165